Using the Calibration Programs 
9-4 
all necessary information that is pertinent to your analyzer. Suggested items that you may 
want to include in the log table are: 
 Calibration date 
  Supplier of calibrator 
 Lot number 
  Expected results and limits 
  Result of background check. 
 
Enter the administrator password as instructed in Chapter 5.2.1 and then choose one or 
several parameters among WBC, RBC, HGB, MCV and PLT for calibration.   
 
9.3.2    Calibration Using Calibrator Program 
 
ENTER the administrator password in the “Password” screen. Press [MENU] to enter the 
system menu.   
 
Figure 9-1 System menu 
SELECT “Calibration→ Calibrator” (Figure 9-1) to enter the “Calibrator” screen (Figure 
9-2).  
 
Figure 9-2 “Calibrator” screen