Using the Calibration Programs
9-4
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are:
Calibration date
Supplier of calibrator
Lot number
Expected results and limits
Result of background check.
Enter the administrator password as instructed in Chapter 5.2.1 and then choose one or
several parameters among WBC, RBC, HGB, MCV and PLT for calibration.
9.3.2 Calibration Using Calibrator Program
ENTER the administrator password in the “Password” screen. Press [MENU] to enter the
system menu.
Figure 9-1 System menu
SELECT Calibration Calibrator” (Figure 9-1) to enter the “Calibrator” screen (Figure
9-2).
Figure 9-2 “Calibrator” screen