Using the Calibration Programs 
9-5 
Selecting the count mode 
Press [F1] to select desired calibration mode. 
 
Editing calibration settings 
Press [F2] to activate the edit boxes, as Figure 9-4 shows. 
 
Figure 9-3 Active “Calibrator” screen 
  Entering lot number 
ENTER the lot number of the calibrator to be used into the “Lot No.” box.   
 
  Entering Exp. Date 
ENTER the expiration date of the calibrator to be used into the “Exp. Date” box.   
 
  Entering the expected results (mean) and limits (range) 
ENTER the expected results (mean) into the “Mean” box of the parameters to be included in 
the calibration.   
 
 
z  Refer to the instructions of use of the calibrator for information on the lot 
number, expiration date, expected results and limits. 
z  Open reagents are stable for 60 days. The entered expiration date should be 
the open date + 60 days or the expiration date marked on the packaging of 
the reagent, whichever is earlier.   
z  When editing the settings, if you want to correct an erroneous entry, 
MODIFY the wrong digit.