Maintaining Your Analyzer
10-21
clean the probe wipe, as Figure 10-29 shows. CLICKYes” to do the procedure; CLICKNo
to abort the procedure.
Figure 10-29 message box
10.3.13 Prepare to Ship
Use the “Prepare to ship” program to prepare your analyzer for a prolonged period of
non-use or for shipping.
Follow the steps given below to do so:
1. Press the appropriate arrow keys ([][][][]) to move the cursor to ”Prepare to
ship”. Remove the diluent, rinse and lyse tubing from the containers following the
instructions displayed on the screen;
2. Press [ENTER] and a message box will pop up to ask you to confirm this operation, as
Figure 10-30 shows;
Figure 10-30 A message box of prepare to ship
3. CLICKNo” if you want to abort this operation; CLICKYes” to proceed with the
operation. The analyzer starts to drain the fluidic lines and the progress is displayed on
the screen, as Figure 10-31 shows.