Customizing the Analyzer Software
5-13
5.3.4 Gain
You can select the “Gain” group to view or change (if you have the administrator password)
the WBC, RBC and HGB gains.
Figure 5-21 Setting gain
Selecting the Gain” group
Press [F1] to select the “Gain” group, as Figure 5-21 shows.
Setting the WBC gain
When WBC histograms of most samples are similar to Figure 5-22, it implies too small a
WBC gain and you need to increase the gain appropriately.
Figure 5-22 WBC gain too small
When WBC histograms of most samples are similar to Figure 5-23, it implies too large a WBC
gain and you need to decrease the gain appropriately.
Figure 5-23 WBC gain too large