Operating Your Analyzer
6-5
6.5 Selecting Analysis Mode
At the “Count” screen, press [MODE] to select one of the six analysis modes. The selected
mode will be displayed in the Count Mode area.
1. Whole - All
It means the sample to be analyzed is a whole blood sample and all the 19 parameters are to
be analyzed.
2. Whole - WBC/HGB
It means the sample to be analyzed is a whole blood sample and only the following
parameters are analyzed: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran% and HGB,
plus the WBC histogram.
3. Whole -RBC/PLT
It means the sample to be analyzed is a whole blood sample and only the following
parameters are analyzed: RBC, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV,
PDW and PCT, plus the RBC and PLT histograms.
4. Prediluted - All
It means the sample to be analyzed is a prediluted blood sample and all the 19 parameters
are to be analyzed, plus 3 histograms.
5. Prediluted - WBC/HGB
It means the sample to be analyzed is a whole blood sample and only the following
parameters are analyzed: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran% and HGB,
plus the WBC histogram.
6. Prediluted -RBC/PLT
It means the sample to be analyzed is a prediluted blood sample and only the following
parameters are analyzed: RBC, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV,
PDW and PCT, plus the RBC and PLT histograms.