Using the Calibration Programs
9-10
Verifying new calibration factors
Press [MENU] to exit the “Calibrator” screen, a message box will pop up to confirm the new
calibration factors, as Figure 9-9 shows.
Figure 9-9 A message box to confirm the new calibration factors
CLICK Yes” to save the new calibration factors and enter the “Count” screen from the
system menu.
At the “Count” screen, run the calibrator or a normal-level control at least 3 consecutive times
and calculate the mean of the results. Compare the obtained means to the expected means.
Other operations
Printing new calibration factors
Press [PRINT] to print out the current calibration factors.
Exiting the ”Calibrator” screen
Press [MENU] to exit to system menu.
9.3.3 Fresh Blood Calibration
ENTER the administrator password in the “Password” screen. Press [MENU] to enter the
system menu.
Figure 9-10 System menu