Using the Calibration Programs
9-9
Figure 9-7 A message box to confirm the validity
CLICK Yes” to save the results; CLICKNo” to abort the result. The saved results will be
displayed on the screen.
Repeat the above steps to run the calibrator 3 to 5 times (5 is recommended) and the
analyzer will automatically calculate the CVs and calibration factors, as Figure 9-8 shows.
Figure 9-8 Results of the calibration
The calculated calibration factor should be within the 75 to 125. If not, there will be
flagged with a “*”. Other values will not be displayed. In case of an empty calibration factor, try
to find out the reason and if necessary, contact Mindray customer service department or your
local distributor for assistance.