Maintaining Your Analyzer
10-37
10.11 Replacing the Filter of the Vacuum Chamber
1. Unscrew and remove the retaining screws with hands or screwdrivers (pointed by the
arrows shown in Figure 10-56 ) on the right plate;
Figure 10-56 Removing the two screws
2. Follow the arrow shown in Figure 10-57 to push and remove the right plate;
Figure 10-57 Removing the right plate
3. Find the filter shows in Figure 10-58;