Using the QC Programs
8-18
Press [PRINT] to print out all the settings.
Exiting the “X-B Edit” screen
Press [MENU] to exit the “X-B Edit” screen. a message box will pop up to remind you to save
the changes, as shows. CLICK
Yes” to save the changes; CLICKNo” to abort the changes.
Figure 8-24 A message box to confirm the changes
8.3.2 Running X-B Analysis
After you have enabled the X-B analysis and assigned valid means and ranges to the three
parameters, the system will automatically run the X-B analysis every 20 patient samples, of
which the MCV, MCH and MCHC results are within the set ranges.
8.3.3 Reviewing X-B Results
This analyzer automatically saves the results of the X-B analyses. You can review the saved
results in two modes - X-B table and X-B graph
X-B Table
Follow the steps introduced in Chapter 8.3.1.
X-B Graph
At the X-B Table” screen, press [F2] to enter the “X-B Graph” screen, as Figure 8-25 shows.