Troubleshooting Your Analyzer
11-2
11.2 Errors without available error messages
Error Possible cause(s) Recommend action
The analyzer
cannot be
turned on.
1. power cord is broken or not
well connected;
2. The fuse is broken;
3. The power outlet has no
electricity.
1. check the power cord connection;
2. Check the fuse;
3. Check the electrical outlet.
Liquid drips
from analyzer
inside.
Damaged pump hose or blocked
filter.
1. Turn off the power and wipe the
analyzer dry;
2. Call Mindray customer service
department or your local distributor
for assistance.
Recorder does
not work.
1. Recorder paper is jammed;
2. Something is wrong with the
circuit.
1. Remove the jammed paper;
2. If the problem remains, turn off the
analyzer and turn it on again in 10
seconds.