Maintaining Your Analyzer
10-6
You should perform the “Rinse prime” procedure to prime the rinse tubing when
there are bubbles in the tubing; or
the rinse in the tubing is contaminated; or
you have installed a new container of rinse without shutting the analyzer.
At the “Maintenance” screen, SELECTRinse prime” to prime the tubing and the priming
progress will be displayed at the bottom of the screen, as Figure 10-4 shows. When the
priming is done, the screen will return to the initial state.
Figure 10-4 Priming rinse screen
10.3.3 Lyse Prime
z Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z After installing new reagents, let the reagents stand for a while before using
them.
z After installing a new container of diluent, rinse or lyse, do a background
check to ensure the background results are normal.