Maintaining Your Analyzer
10-11
Figure 10-11 A message box to confirm the cleaning
10.3.7 E-Z Cleanser Cleaning
z Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
You can use the E-Z cleanser, an enzyme based, isotonic cleaning solution and wetting agent,
to clean the tubing and bath by performing the “E-Z cleanser cleaningprocedure.
Follow the steps given below to perform the procedure:
1. At the “Maintenance” screen, move the item to “E-Z cleanser cleaning”;
2. Present the cleanser to the probe and press [ENTER] to aspirate the cleanser. When you
hear the beep and the sample probe is out of the bottle, remove the cleanser. This
analyzer will automatically prime the bath and fluidic lines with the aspirated cleanser and
the progress is displayed on the screen, as Figure 10-12 shows;
Figure 10-12 Priming the bath and fluidic lines
3. When the priming is done, the analyzer will start the 10-minute soaking process, as