Customizing the Analyzer Software
5-5
This area displays useful information to help you move to the next step.
At this screen, if you want to acquire help information, press [HELP]; if you want to return to
the system menu, press [MENU].
5.3.1 Reagent
You can select the “Reagent” group to change the settings regarding the reagents and the
waste, as Figure 5-7 shows.
Figure 5-7 “Reagent” settings
Selecting the “Reagent” group
Press [F1] to select the “Reagent” group.
Setting remaining volumes for reagents
You may set the remaining volumes for the diluent, rinse and lyse. When any of the entered
volumes is counted down to zero, the system will remind you to install a new container.
1. SELECT “Diluent”, “Rinse” or “Lyse” in the “Remains” field as needed;
2. ENTER the desired digits. See Table 5-1 for the valid reagent volumes.
Table 5-1 Valid reagent volumes
Diluent Rinse Lyse
Allowed range
About 0 to 30.0 L About 0 to 30.0 L About 0 to 999.0 mL
Entering the usable volume of the waste container
You may enter the usable volume of the waste container. When the system counts down the
entered volume to 0, it will alert you to empty the waste container. Follow the steps given
below to set the volume.