Customizing the Analyzer Software
5-14
To increase (or decrease) the gain
1. Enter the administrator password as introduced in Chapter 5.2.1.
2. At the ”Gain” screen and ENTER the desired gain into the “WBC (WB), as Figure 5-24
shows, or “WBC (PB)”, as Figure 5-25 shows.
Figure 5-24 Setting WBC (WB) gain
Figure 5-25 Setting WBC (PB) gain
Setting the RBC gain
When the difference between the actual MCV result and the expected result exceeds 6%,
you need to change the RBC gain.