Maintaining Your Analyzer
10-17
Figure 10-22 The time of draining bath
4. Check the bath and the tubing below them for residual fluid. If there is no fluid, press
[ENTER] to prime the bath with diluent, as Figure 10-23 shows. When the priming is done,
the screen will return to the initial state;
Figure 10-23 Priming the bath with diluent
5. If there is fluid left, turn off the analyzer and call Mindray customer service department or
your local distributor for assistance.