Troubleshooting Your Analyzer
11-4
5.3.4 to set the voltage within 3.4 to
4.8V, preferably 4.5V;
3. If the problem remains, shut down
your analyzer and contact Mindray
customer service department or you
local distributor.
HGB adjust HGB blank voltage within 3.2
V to 3.4 V or 4.8 V to 4.9 V.
1. Do the probe cleanser cleaning
procedure as instructed in Chapter
10.3.6.;
2. If the problem remains, adjust the
HGB gain as instructed by Chapter
5.3.4 to set the voltage within 3.4 to
4.8V, preferably 4.5V;
3. If the problem remains, shut down
your analyzer and contact Mindray
customer service department or you
local distributor.
WBC clog
1. Clogged the aperture;
2. Inappropriate WBC count
time setting;
3. Solenoid valve error.
1. Enter the “Service Maintenance
screen. Zap and flush the aperture as
instructed by Chapter 10.3.4 and
10.3.5;
2. Enter the “Setup Settings
Count” screen and record the WBC
count time. Then enter the “Service
Self-test” screen and test the actual
WBC count time as instructed by
Chapter 10.5.1.;
3. If the difference between the reference
WBC count time and the actual WBC
count time is less than 2 seconds, the
error has been removed;
4. If not, enter the “Service
Maintenance” screen and do the
probe cleanser cleaning procedure as
instructed by Chapter 10.3.6;
5. Enter the “Setup Settings
Count” screen and record the WBC
count time. Then enter the “Service
Self Test” screen and test the actual
WBC count time as instructed by
Chapter 10.5.1;