Troubleshooting Your Analyzer
11-11
Reagents” and adjust the remaining
diluent volume as instructed by
Chapter 5.3.1;
2. If not, change a new container of rinse
as instructed by Chapter 4.4.1.
Waste full
The waste container is full. Empty the waste container and reset
usable volume of the waste container as
instructed by Chapter 5.3.1.
File error
Something is wrong with the
analyzer’s file system.
Turn off the power and contact Mindray
customer service department or your local
distributor.
Dynamic
memory error
Something is wrong with the
analyzer’s memory.
Turn off the power and contact Mindray
customer service department or your local
distributor.
+12 V power
error
Something is wrong with the
power board.
Shut down the analyzer and contact
Mindray customer service department or
your local distributor.
-12 V power
error
Something is wrong with the
power board.
Shut down the analyzer and contact
Mindray customer service department or
your local distributor.
56V power error Something is wrong with the
power board.
Shut down the analyzer and contact
Mindray customer service department or
your local distributor.