Mindray BC-2800 - 6.2.11 Draining Tubing Procedure

Mindray BC-2800
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Customizing the Analyzer Software
5-18
*.** 10
12
/L Default
*.** 10
6
/uL /
*** 10
4
/uL /
RBC
*.** /pL /
**.* % Default HCT
.*** L/L /
***.* fL Default MCV, RDW-SD
***.* um
3
/
RDW-CV **.* % Default
**** 10
9
/L Default
**** 10
3
/uL /
***.* 10
4
/uL /
PLT
**** /nL /
***.* fL Default MPV
***.* um
3
/
PDW **.* / Default
.*** % Default PCT
*.** mL/L /
*.*** pg Default
MCH
**.** fmol /
Follow the steps given below to select the units.
1. Enter the administrator password as introduced in Chapter 5.2.1;
2. At the “Settings” screen, press [F1] to select the “Count” group;
3. SELECT the desired unit from the drop-down list of the desired parameter, as Figure
5-30 shows.