Customizing the Analyzer Software
5-22
Setting reference ranges (administrators only)
1. Enter the administrator password as introduced in Chapter 5.2.1;
2. At the “Settings” screen, press [F1] to select the “Ref. Range” group, as Figure 5-35
shows;
3. When you have finished selecting the patient group, SELECT the desired parameter and
ENTER the desired digits for the upper and lower limits;
Figure 5-35 Setting reference range
4. Press [F2] to save the changes. If the changes are successfully saved, a message box
shown in Figure 5-36 will pop up; CLICK the “Yes” to close the message box;
Figure 5-36 A message box to note data saved
5. If some entered limits are invalid, a message box shown in Figure 5-37 will pop up.
CLICK Yes” and enter valid number again;