3-7 
 
You can either perform a manual leakage test: 
1.  Perform procedures 1-4 in the NIBP Accuracy Test section. 
2.  Raise the pressure in the rigid vessel to 250 mmHg with the balloon pump. Then, wait 
for 5 seconds to let the measured values becoming stable. 
3.  Record the current pressure value and meanwhile use a time counter to count time. Then, 
record the pressure value after counting to 60s. 
4.  Compare the two values and make sure the difference should not be greater than 6 
mmHg. 
 
 
3.2.3 Sidestream and Microstream CO
2
 Module Tests 
Leakage test 
Follow this procedure to perform the test: 
1.  Plug the module into the module rack. 
2. Wait until CO
2
 warmup is finished and then use your hand or other objects to 
completely block the gas inlet of the module or watertrap. The sidestream and 
microstream CO
2
 modules will behave as follows: 
Sidestream: The alarm message [CO
2
 FilterLine Err] is displayed on the screen after 3 
seconds. Block the gas inlet for another 60 s. Select [Main Menu]→ 
[Maintenance >>]→[User Maintenance >>] → enter the required password  →
[Maintain CO
2 
>>]→ [Calibrate CO
2 
>>], and check that the flow rate is less than. 
10ml/min. The module does not leak if current flow rate is less than 10ml/min and 
the alarm message does not disappear.
 
 
Microstream: The alarm message [CO
2
 Purging] is displayed on the screen after 
certain time. Block the gas inlet for another 30s. If alarm message [CO
2
 FilterLine 
Err] is shown, it indicates that the module does not leak.