Nurse call relay performance test
The relay contacts are close when an alarm occurs. Yes No
Analog output performance test
The waves displayed on the oscillograph are identical with those
displayed on the monitor.
Yes N o
Electrical safety tests
Refer to A Electrical Safety Inspection. All the electrical safety
tests should be passed.
Yes N o
Touchscreen calibration
The touchscreen is calibrated successfully. Yes No
Recorder check
The recorder can print ECG waves correctly and the printout is
clear.
Yes N o
Set the recorder to some problems such as out of paper, paper
jam, etc. the monitor gives corresponding prompt messages.
After the problem is removed, the recorder is able to work
correctly.
Yes N o
Automatic alarm recording for each parameter functions
correctly when parameter alarms occur.
Yes N o
Network print test
The network printer can print out ECG reports correctly. Yes No
Battery check
The monitor can operates correctly from battery power when an
AC power failure accidentally occurs.
Yes N o
T5 patient monitor can operate independently on a single battery. Yes No
The operating time of the battery meets the product
specification.
Yes N o
Test conclusion:
Tested by:
Test Date: