Structure and Assembly/Disassembly 9-9
4. Unscrew two M3 step screws fixing PCBA, and remove fan PCBA.
5. Pull out of the fan socket to remove the fan. See the figure.
9.3.5 Disassembling Battery Assembly
The disassembly tool: cross-headed screwdriver (M3, M4), diagonal cutting pliers, anti-electrostatic
glove.
1. Follow step 1 and step 2 in Chapter 9.3.2 to remove left and right side panels.
2. Unscrew two M4 X 8 screws fixing the rear cover, and remove the rear cover.