Adjustments 7-7
7.1.4 Monitor Parameter Setting
After changing the main monitor, parameter loading should be performed to match the
monitor and the system.
The parameters of the main monitor include color temperature. Information of loading Default
parameters are introduced here:
1. Press <1> and <4> key on the monitor as shown in the figure for more than 3 seconds to enter
the screen.
2. Press <1> or <2> to select Default, the cursor is on the Default position, press <4> to enter the
parameter automated loading, and it takes about 3 seconds to finish.
3. Color temperature is default to be 11000K, and adjustment is not recommended.
4. After all settings are finished, move the cursor to Exit or press <3> to exit the menu.
7.2 Touch Screen Adjustment
7.2.1 Touch Screen Brightness and Contrast Adjustment
Avoid operating under direct sunlight, or the touch screen could be blocked. Do not
place anything on the touch screen, or the screen view will be blocked.
Open the screen through the path: [Setup]->[System]->[General], and click [Touch Screen Preset]
to perform the adjustment.