C-18 Description of Self Test Items
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace mini CW board if the test result is FAIL.
C.1.36 Z1301 DSP FPGA and 4D&TEE Board
Interconnection Test
Top test item
PC module and DSP FPGA interconnection test
Test content
The program checks 4D probe is not plugged in, and then read the signal in the place to decide
whether 4D&TEE board is in the place, then read 4D&TEE board’s ID. The interconnection
between DSP FPGA and 4D&TEE board works well if it appears non-0 and non-F.
Analysis to test failure
The drive goes wrong if the test result appears Error.
Mini 4D&TEE board is not in the place if the information “4D&TEE board is not in the place”
appears or the circuit is not in the place.
The interconnection between DSP FPGA and 4D&TEE board goes wrong if the test fails.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace 4D&TEE board if the test result is FAIL.
C.1.37 Z1302 4D&TEE Board Driver Circuit Test
Top test item
Communication test between DSP FPGA and 4D&TEE board
Test content
Set DA output sine signals of Sin+ drive circuit and Sin- drive circuit. After being magnified, AD
collects the signal to decide the frequency and the amplitude. Set output cosine signals of Cos+
drive circuit and Cos- drive circuit. After being magnified, AD collects the signal to decide the
amplitude.
Analysis to test failure
The drive goes wrong if the test result appears Error.
The drive circuit goes wrong if the information “The test of XXX drive circuit failsappears.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace 4D&TEE board if the test result is FAIL.