System Installation 3-19
3) Click [Add] to add the server to the list if the connection works normally.
The following is an example:
NOTE:
1.
AE Title should be the same with the SCU AE Title preset in the server
(PACS/RIS/HIS).
2.
DICOM communication port should be the same with the one in the server.
3.
If the currently entered name has already existed, the system will pop up: “The
server name exists!” Click [OK] to enter another name.
3. Click [Set DICOM Service].