Passport 12/Passport 8 Operators Manual A-19
Alarm limit Range Step
EtCO
2
High (low limit + 2) to 99 mmHg
1 mmHg EtCO
2
Low 1to (high limit - 2)mmHg
FiCO
2
High 1 to 99 mmHg
awRR High
Adult, pediatric:
Neonate:
(low limit + 2) to 100 rpm
(low limit + 2) to 150 rpm
1 rpm
awRR Low 0 to (high limit - 2)rpm
EtO
2
High (low limit + 2) to 100 %
0.1%
EtO
2
Low 18 to (high limit - 2)%
FiO
2
High (low limit + 2) to 100 %
FiO
2
Low 18 to (high limit - 2)%
EtN
2
O High (low limit + 2) to 100 %
1%
EtN
2
O Low 0 to (high limit - 2)%
FiN
2
O High (low limit + 2) to 100 %
FiN
2
O Low 0 to (high limit - 2)%
EtHal/Enf/Iso High (low limit + 0.2) to 5.0 %
0.1%
EtHal/Enf/Iso Low 0 to (high limit - 0.2)%
FiHal/Enf/Iso High (low limit + 0.2) to 5.0 %
FiHal/Enf/Iso Low 0 to (high limit - 0.2)%
EtSev High (low limit + 0.2) to 8.0 %
0.1%
EtSev Low 0 to (high limit - 0.2)%
FiSev High (low limit + 0.2) to 8.0 %
FiSev Low 0 to (high limit - 0.2)%
EtDes High (low limit + 0.2) to 18.0 %
0.1%
EtDes Low 0 to (high limit - 0.2)%
FiDes High (low limit + 0.2) to 18.0 %
FiDes Low 0 to (high limit - 0.2)%