C-6 Passport 12/Passport 8 Operator’s Manual
C.1.4 SpO
2
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes SpO
2:
:
Desat:
Med
High
SpO
2
High Yes Yes When used with a DPM central station
Adu, Ped:
Neo:
100
95
When used with a Panorama central
station
Adu, Ped:
Neo:
Off
95
SpO
2
Low Yes Yes 90
Desat Yes Yes 80
Sensitivity (Masimo) Yes Yes Normal
Averaging (Masimo) Yes Yes 8 s
Sat-Seconds (Nellcor) Yes Yes 0 s
Sweep Yes Yes 25 mm/s
NIBP Simul No Yes Off
C.1.5 Temp
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
T1/T2 High (ºC) Yes Yes 38.0
T1/T2 Low (ºC) Yes Yes 35.0
TD High (ºC) Yes Yes 2.0

Table of Contents