Passport 12/Passport 8 Operators Manual C-7
C.1.6 NIBP
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
Interval Yes Yes
General: 15 min
OR: 5 min
ICU: 15 min
NICU: 30 min
CCU: 15 min
NIBP End Tone Yes Yes Off
Clock Yes Yes On
Initial Pressure (mmHg) Yes Yes
Adu:
Ped:
Neo:
160
140
90
Cuff Press. (mmHg) Yes Yes
Adu:
Ped:
Neo:
80
60
40
Alarm Limit
NIBP-S High (mmHg) Yes Yes
Adu:
Ped:
Neo:
160
120
90
NIBP-S Low (mmHg) Yes Yes
Adu:
Ped:
Neo:
90
70
40
NIBP-M High (mmHg) Yes Yes
Adu:
Ped:
Neo:
110
90
70
NIBP-M Low (mmHg) Yes Yes
Adu:
Ped:
Neo:
60
50
25
NIBP-D High (mmHg) Yes Yes
Adu:
Ped:
Neo:
90
70
60
NIBP-D Low (mmHg) Yes Yes
Adu:
Ped:
Neo:
50
40
20