IBP Monitoring
15-7
IBP alarm limits
Pressure Label Max. Upper (mmHg) Min. Lower (mmHg) Step (mmHg)
ART 300 0 1
PA 120 -6 1
CVP 40 -10 1
RAP 40 -10 1
LAP 40 -10 1
ICP 40 -10 1
P1 300 -50 1
P2 300 -50 1
SCALE
ADJUST >>
This option allows you to open the following menu and set
the positions of the high, reference and low scales for the
two waveforms.
Figure 15-7 IBP Scale Adjust Menu
1. HI: Determines the high scale. The reference scale changes when the high scale
is changed.
2. LO: Determines the low scale. The reference scale changes when the low scale
is changed.
3. VAL: Determines the reference scale and position. The high and low scales
remain unchanged when the reference scale is changed.
NOTE
z The high scale should not be smaller than the low scale; the low scale
should not be larger than the high scale.
z The waveform amplitude changes when the high and low scales are
adjusted.