Appendices
20-33
C.2.9 IBP Module Alarm Messages
Alarm message A B L Cause Measure
IBP1 SENSOR OFF No Yes 3 The invasive blood pressure
cable of channel 1 is
disconnected from the monitor.
Check if the IBP 1 sensor is
properly connected.
IBP2 SENSOR OFF No Yes 3 The invasive blood pressure
cable of channel 2 is
disconnected from the monitor.
Check if the IBP 2 sensor is
properly connected.
IBP1 NEED
ZERO-CAL
No No 3 The IBP transducer of channel
1 has not been zeroed.
Zero the IBP transducer of channel
1.
IBP2 NEED
ZERO-CAL
No No 3 The IBP transducer of channel
2 has not been zeroed.
Zero the IBP transducer of channel
2.
C.2.10 CO Module Alarm Messages
Alarm message A B L Cause Measure
TB SENSOR OFF No Yes 3 The cable for blood pressure
measurement is disconnected from
the monitor.
Check for proper connection
of TB cable.
C.2.11 Mindray CO
2
Module Alarm Messages
Alarm message A B L Cause Measure
CO2 NO
WATERTRAP
No No 2 The water trap is not properly
connected or is disconnected.
Make sure the sidestream CO
2
water trap is firmly connected.
CO2 SENSOR
TEMP HIGH
No No 1 The temperature of the sensor
assembly is too high.
CO2 SENSOR
TEMP LOW
No No 1 The temperature of the sensor
assembly is too low.
CO2 AIRWAY
PRES TOO HIGH
No No 2 The pressure inside the airway
is too high.
CO2 AIRWAY
PRES TOO LOW
No No 2 The pressure inside the airway
is too low.
CO2
BAROMETRIC
TOO HIGH
No No 2 The barometric pressure is too
high.
Restart the monitor if necessary. If
the problem remains, contact our
company for repair.