Appendices
20-37
C.2.14 AG Module Alarm Messages
Alarm message A B L Cause Measure
AG SENSOR OFF No Yes 3 The AG sensor is not
properly connected or is
disconnected.
Make sure the AG sensor is
properly connected.
AG NO WATERTRAP No Yes 3 The AG water trap is
disconnected from the
monitor.
Make sure the water trap is
properly connected.
CHANGE AG WATERTRAP Yes No 2 The AG water trap is
replaced.
Wait until the water trap
has been changed.
AG WATERTRAP TYPE
WRONG
Yes No 2 The AG water trap is of
a wrong type.
Replace with a water trap
of a correct type.
AG USA ERROR Yes No 2 AG module fault. Contact Mindray to get rid
of the fault.
AG PARAMAGNETIC O2
ERR
Yes No 1
AG GALVANIC O2 ERROR Yes No 1
AG OXIMA DEPLETION
WARN
Yes No 2
AG OXIMA DEPLETION
ERR
Yes No 2
O
2
module fault. Contact Mindray to get rid
of the fault.
AG OCCLUSION Yes No 1 The actual pump rate of
the AG module is
<20ml/min, which
exceeds 1 second.
Remove the block of the
airway.
AG HARDWARE ERROR Yes No 1 Hardware fault of the
AG module.
AG DATA LIMIT ERROR Yes No 2 AG module fault.
AG ZREF FAIL Yes No 3 Failure in AG module
zeroing.
AG CAL FAIL Yes No 1 Failure in AG module
calibration.
Contact Mindray to get rid
of the fault.
C.2.15 Recorder Module Alarm Messages
Alarm message A B L Cause Measure
RECORDER INIT
ERR N
Yes No 2 An error occurs during the
recorder initialization.
Contact the hospital’s engineers or
our customer Service.
Note: N represents the error number.