Appendices
20-38
Alarm message A B L Cause Measure
RECORDER
SELFTEST ERR
Yes No 2 An error might occur to the
RAM, ROM and CPU
watchdog.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.
RECORDER VLT
HIGH
No No 1
RECORDER VLT
LOW
No No 1
A problem occurs to the system
power.
If this alarm message is given for
many times, contact our company
for repair.
RECORDER
HEAD HOT
No No 1 The thermal head of the
recorder is too hot.
Resume the recording till the
recorder cools down completely. If
the problem still exists, contact our
company for repair.
REC HEAD IN
WRONG POS.
Yes Yes 3 The thermal head of the
recorder is in wrong position.
Restore the control lever of the
recorder to its previous position.
RECORDER OUT
OF PAPER
Yes Yes 3 The recorder paper is used up. Replace with a new paper roll.
RECORDER
PAPER JAM
No No 2 The recording continues for
more than 30 minutes.
Place the recorder correctly and
try again.
RECORDER
COMM ERR
Yes No 2 Error in recorder
communication.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.
TOO MANY REC
TASKS
No No 2 Quite a few alarm events occur
at the same time.
Check the patient’s condition and
the alarms. Open the RECORD
menu and select the CLEAR REC
TASK option. If the problem
remains, contact our company for
repair.
RECORDER
PAPER W.P.
Yes Yes 2 The paper roll of the recorder is
not placed in the correct
position.
Place the paper roll correctly.
RECORDER S.
COMM ERR
Yes No 2 Error in recorder
communication.
REC NOT
AVAILABLE
No No 2 Error in the recorder work
mode.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.