Appendices
20-41
Prompt messages Cause Measure
UP startup. warming itself up.
CO2 CALIBRATE The CO
2
module is in the calibrating status. Wait for the CO
2
module to finish
the calibration.
CO2 PURGING The CO
2
module is in the purging status. Wait for the CO
2
module to finish
the purging.
Recorder
RECORDER
INITIALIZING
The recorder is in the initializing status. Wait for the recorder to finish the
initialization.
RECORDER BUSY The recorder is recording. Wait for the recorder to finish the
recording.
AG module
AG STANDBY The AG module is in the standby mode. None
AG IS STARTING The AG module is starting. Wait for the AG module to finish
the startup.
AG WARM UP The AG module is warming itself up. Wait for the AG module to finish
warming itself up.
NIBP module
Manual measure... The NIBP module is performing the manual
measurement.
CONTINUAL... The NIBP module is performing the
continuous measurement.
Auto measuring... The NIBP module is performing the auto
measurement.
Wait for the NIBP module to finish
the measurement.
Resetting...
Resetting...
The NIBP module is being reset. Wait the NIBP module to finish
the resetting.
Please start This message appears after the auto
measurement interval is selected.
Press the NIBP button to start the
measurement.
CALIBRATE... The NIBP module is performing the
calibration.
Wait for the NIBP module to finish
the calibration.
Calibration over The calibration is finished. None
PNEUMATIC... The NIBP module is checking the pneumatic
system for leakage.
Wait for the NIBP module to finish
checking the pneumatic system.
Pneum test over The NIBP finishes checking the pneumatic
system for leakage.
None