C.3 Prompt Messages
Prompt messages Cause Measure
ECG1 SIGNAL
SATURATION
ECG2 SIGNAL
SATURATION
Signals of abrupt change interfere with the
ECG signal.
Check whether the electrodes and
leads are well connected.
SEARCH PULSE The SpO
2
module is searching the pulse. Wait till the end of the searching.
Welch Allyn CO2 module
CO2 STANDBY The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 WARM UP
The CO
2
module is starting and warming
itself up.
Wait for the CO
2
module to finish
warming itself up.
CO2 SENSOR START
UP
The CO
2
sensor is starting.
Wait for the CO
2
sensor to finish
the startup.
Mindray CO2 module
CO2 STANDBY The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 START UP The CO
2
module is starting. Wait for the CO
2
module to finish
the startup.
CO2 CALIBRATE
ZERO
The CO
2
module is in the zeroing status. Wait for the CO
2
module to finish
the zeroing.
CO2 WARM UP The CO
2
module is warming itself up after
startup.
Wait for the CO
2
module to finish
warming itself up.
Oridion CO
2
module
CO2 STANDBY The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 START UP The CO
2
module is starting. Wait for the CO
2
module to finish
the startup.
CO2 CALIBRATE
ZERO
The CO
2
module is in the zeroing status. Wait for the CO
2
module to finish
the zeroing.
CO2 SENSOR START The CO
2
sensor is warming itself up after Wait for the CO
2
module to finish