Setup 12-17
12.10.4 Probe Check
This function enables users to check if a transducer element is in malfunction, so as to evaluate the
transducer performance.
Click [Probe Check] to open the Probe Check screen. The system automatically performs probe
check to the element of the currently activated probe. If a transducer element is in malfunction, it is
displayed as a red spot. If a transducer element functions well, it is displayed as a green spot.
You can export the result image to the external device.
Notes:
Each time a transducer is connected to the main system or activated, the users are
recommended to implement probe check of transducer performance.
Before probe check, make sure the probe is clean and not being used for scan.
12.10.5 Other Settings
If you have any questions, please contact the service engineer or your agent.
12.11 Security
Click [Security] on the Setup menu to enter the security-setting screen.
1. Drive Encryption/Secure Data Wipe
Setup
Export Log Export the log.
Battery Manufacture
State
Displays the battery state.
Self-Tes t Perform system self-test and restart the machine.
Recover
Recover the ultrasound system (including Operating System and
Doppler).
Enter Windows
For this function, you need a one-time password, please contact the
service engineer or your agent.
Prepay Installment Display the prepay installment information.
Start/Stop WinProxy Cursor debug interface.
Fast Startup
To speed up the startup process.
Note: After installing McAfee, please disable the Fast Startup
function to avoid McAfee failure.