Probes and Biopsy 13-47
Relative humidity: 20% to 95% (no condensation)
Store or transport the needle-guided bracket NGB-024 under the following ambient conditions:
Ambient temperature: -10°C to 60°C
Relative humidity: 5% to 95% (no condensation)
13.2.10 Disposal
Be sure to dispose the needle-guided bracket only after sterilizing it.
Contact your MINDRAY representative when disposing of this device.
13.3 Needle Navigation Guiding
WARNING
The person performing needle navigation guiding
procedures must understand diagnostic ultrasound
thoroughly and have been trained adequately,
otherwise, side effects may be caused in the patient.
Wear sterile gloves before performing the needle
navigation.
Clean and disinfect the navigation device by instructions
in the device accompanied manual before and after
needle navigation.
Please read the warning information by instructions in
the device before needle navigation.
Disinfect the probe and sterilize the bracket. For details,
refer to chapter 13.2.8 Clean and Sterilize the Needle-
guided Bracket”. Otherwise, cross-infection may be
caused.
When performing needle navigation procedures, use
only sterile ultrasound gel that is certified to be safe.
And manage the ultrasound gel properly to ensure that
it does not become a source of infection.
Needle navigation on patients with implants or with
intracavity electro, magnetic, mechanical device are
prohibited. When the magnetic navigator is working,
previously described patients should be at least 1
meter away from the working area. Bent or damaged
needle are forbidden in the operation.
Perform accuracy verification before conducting needle
navigation. You can use the needle-guided bracket to
anchor the needle before needle navigation for a better
result.