System Preparation 3-1
3 System Preparation
WARNING:
1.
Do not connect the three-wire cable of the system with a two-wire
plug without protective grounding; otherwise, electric shock may
result.
2.
Do connect the power plug of this system to wall receptacles that
meet the ratings indicated on the rating nameplate. If adapters or
multifunctional receptacles are used, it may cause the leakage
current to exceed the safety requirement.
3.
In the environment that patient is 1.5 meters around, connect
peripherals to the auxiliary power outlet which is capable of
isolation protection, or power the peripherals by auxiliary output
cable or isolation transformer complied with
IEC60601-1 or the
power input of the same safety level.
4.
DO NOT use power supply of different phases to power peripherals,
like power supply of air-conditioning.
5.
When using peripherals not powered by the auxiliary output of the
ultrasound system, or using peripherals other than permitted by
Mindray, make sure the overall leakage current of peripherals and
the ultrasound system meets the requirement of the local medical
device electrical regulation (like enclosure leakage current should
be no more than 500uA of IEC60601-1), and the responsibility is held
by the user.
3.1 Moving/Position the System
Please read and understand the safety precautions before placing the system to ensure safety for both
operator and devices.
If the ultrasound system is configured with the central brake system:
1. Switch off the power, and pull out the plug.
2. Disconnect all cables from the off-board peripheral devices (printer, recorder, etc.)
3. Unlock the central brake pedal and reset the green pedal of the double-lock caster; hold the handle
to move the system.
4. When you move the system to a desired location, press down the central brake pedal and red
pedal of the double-lock caster to lock the four casters.
If the ultrasound system is not configured with the central brake system:
1. Switch off the power, and pull out the plug.
2. Disconnect all cables from the off-board peripheral devices (printer, recorder, etc.)
3. Unlock the four casters; hold the handle to move the system.
4. When you move the system to a desired location, lock the four casters.