EasyManua.ls Logo

Mindray Recho R9 - Page 348

Mindray Recho R9
477 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13-28 Probes and Biopsy
<9> Specification of guiding block Matched with the corresponding biopsy needle.
<10> Mark of indicating scales Indicating needle distance scales.
<11>
Knob of fixing the guiding
block
Used for fixing the guiding block.
NGB-053/NGB-054
13.2.2 Basic Procedures for Biopsy Guiding
1. Select the right needle-guided bracket and needle, install them properly. For details, please refer to
Chapter 13.2.3 Needle-guided Bracket Inspection and Installation.
2. Verify the biopsy guide line.
3. Press <Biopsy> to enter the biopsy. Before biopsy guiding, you can preset bracket model, biopsy
display and guide line dot type.
Tips:
If the current probe has no matched bracket; or, the image is frozen and the guide line is off
before image frozen, then you cant enter Biopsy menu.
4. Select the bracket and guide line according to actual situation.
5. On the Biopsy menu, click [Verify] to enter the Verify menu to verify the guide line. After the
verification, click [Save] on the menu to save the parameter setting. And then click [Exit] to return to
Biopsy menu.
Tips:
Angle locking nut
Angle adjusting block
Needle-shaped
adjusting nut
Locking nuts of the
guided bracket at the
right and left sides
Torsion spring of the
needle pressure cover
Biopsy needle pressure
position
Angle adjusting base
Left clamp of the
bracket
Right clamp of the
bracket
Needle clamping cover

Table of Contents

Related product manuals