13-30 Probes and Biopsy
(4) Adjust the dial scale to the required needle type shift, and then screw the needle fixing nut to
lock the dial scale. (To adjust the dial scale you have to loose the needle fixing nut first.)
(5) Pull the lock pin and close the V-shaped cover to fix the lock pin in the groove of the needle
type adjusting base, so as to install the needle into the guiding hole.
NGB-025
1. Put on the sterile probe sheath.
2. Open the clamp. Insert the front clamp to the front groove.
3. Push the biopsy forward (arrow’s direction) until the locating pole inserting into the location hole.
Turn the lower clamp against the intra-cavity probe. Tighten the nut to lock the biopsy (arrow’s
direction).
NGB-029
1. Put on the sterile probe sheath.
2. Select a proper needle-guided bracket, and match the locating groove with the tab of the probe.
Mount the bracket onto the probe.
Align front clamp with
Insert the locating pole