13-42 Probes and Biopsy
NGB-022
(1) Pull the lock pin and open up the V-shaped cover to expose the needle.
(2) Separate the bracket and the transducer from the needle.
(3) Screw the pinch nut to release the needle-guided bracket.
(4) Separate the bracket and the transducer.
NGB-025
Hold the prober in the left hand; unscrew the locking nut with the right hand to loose the clamp
(arrows direction).Lift the biopsy up (towards arrows direction). The locating pole, front clamp, the
locating hole and the front clamp become loose.
NGB-029
1. Hold the transducer. Press the biopsy needle to separate the needle from pressure position of the
needle.
2. Separate the bracket and the transducer from the needle.
3. Rotate the clamping nuts of the needle guided-bracket (following the direction of the arrow). The
needle guided-bracket is separate from the transducer. Hold the transducer and take out the
bracket.