16-2 EMC Guidance and Manufacturer’s Declaration
11. Portable and mobile RF communications equipment could affect
system. See below tables.
12. Resona I9 series should be away from RFID, MRI, diathermy, and
electrocautery testing, wireless power transfer, 5G cellular and
security equipment (such as electromagnetic anti-theft system
and metal detector). If the devices are near and are interfered by
the concealed and undiscovered RF transmitter (for example,
scanning mode changes or image disturbances affecting
diagnosis), the user should immediately take mitigation measures,
such as redirecting, repositioning or shielding the RF transmitter.
13. This device complies with part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) This device may not
cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause
undesired operation.
14. This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to
radio or television recept
ion, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and receiver.
–
Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
–
Consult the dealer or an experienced radio/TV technician for
help. Any Changes or modifications not expressly approved by the
party responsible for compliance could void the user's authority to
operate the equipment.
The device has been evaluated to meet general RF exposure
requirement.
The device can be used in portable exposure condition without
restriction.
If Resona I9 series is operated within the electromagnetic environment listed in Table 3, Table 4, Table
5, Table 6, Table 7 Resona I9 series will remain safe and will provide the following basic performances:
Imaging;
Doppler acoustic spectral displaying;
Taking measurements;
Patient information;
Date/time information.