16-4 EMC Guidance and Manufacturers Declaration
TABLE 3
GUIDANCE AND MINDRAY DECLARATIONELECTROMAGNETIC IMMUNITY
Resona I9 series is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIROMENT
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact;
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact;
±2 kV, ±4 kV, ±8
kV,±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power supply
lines;
±1 kV for input/output
lines
±2 kV for power supply
lines;
±1 kV for input/output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±0,5 kV, ±1 kV line(s)
to line(s);
±0,5 kV, ±1 kV, ±2 kV
line(s) to earth
±0,5 kV, ±1 kV line(s)
to line(s);
±0,5 kV, ±1 kV, ±2 kV
line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips,
Short
interruptions and
voltage variation
on power supply
input voltage
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
Mains power quality
should be that of a typical
commercial or hospital
environment. If you require
continued operation during
power mains interruptions,
it is recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.