Spectrum
®
/Spectrum OR
Service Manual 0070-10-0556-02 4 - 11
Calibration Procedure Diagnostics
4.4.6.3 Motor Pump Test
The purpose of this test is to determine if the output of the pump is adequate.
FIGURE 4-12 Motor Pump Test
1. Connect the dummy cuff / test chamber to the side panel fitting.
2. Select the Motor Pump Test using the Navigator
control knob to activate the test.
3. Target pressure of 300 mmHg will be displayed on the screen. The time required to
pump to 300 mmHg will also be displayed on screen.
Specifications Pump to 300 mmHg in < 35.0 seconds. – 500 cc test
chamber
Specifications: Pump to 300 mmHg in < 49.0 seconds – 700 cc test
chamber