Diagnostics Calibration Procedure
4 - 10 0070-10-0556-02 Spectrum
®
/Spectrum OR
™
Service Manual
6. Adjust R81 potentiometer and verify the linearity accordingly. See Figure 4-11.
7. Reassemble unit and verify the following static pressure points to ensure proper
calibration.
FIGURE 4-11 Static Calibration screen
Specifications 150 mmHg +/- .1%
Specifications 30 mmHg,100 mmHg, 150 mmHg, 190 mmHg and
250 mmHg. +/-3 mmHg
150 mmHg
Digital Mercury Manometer
Bulb
693cc
Test Chamber
P/N 0138-00-0001-01
T - Fitting
Overpressure Voltage Setpoint
Static Pressure Calibration
Motor Pump Test
Leak Test
Dump Test
Bleed Rate Test
Overpressure Test
Abort Test
Previous Menu
NIBP Test
Select to return to previous menu.
Transducer: 150.0 mmHg
Static Calibration Setup
Bulb
Valve
VR 81
NIBP Module
Adjust
VR 81
Here
NIBP Front
Detail View
Manually
Squeeze
Bulb
Static Pressure Calibration