Diagnostics Calibration Procedure
4 - 14 0070-10-0556-02 Spectrum
®
/Spectrum OR
Service Manual
4.4.6.6 Bleed Rate Test
The purpose of this test is to verify the valves, that control the bleed rate, are functioning
properly.
FIGURE 4-15 Bleed Rate Test
1. Connect the dummy cuff / test chamber to the side panel fitting.
2. Select the Bleed Rate using the Navigator
knob to activate the test.
3. The chamber will inflate to 220 mmHg of pressure. The bleed rate valve will open and
deflate the pressure for 10 seconds. The result of the test will be displayed on the screen.
Specifications: Bleed Rate = 6.0mmHg / sec ± 20% - 500cc test
chamber
Specifications: Bleed Rate = 6.0mmHg / sec ± 20% - 700cc test
chamber