13 Probes and Biopsy
Operator’s Manual 13 - 39
b. Open the clamp. Insert the front clamp to the front groove, and align front clamp with
front groove.
c. Push the biopsy forward (arrow’s direction) until the locating pole inserting into the
location hole. Turn the lower clamp against the intra-cavity probe. Tighten the nut to lock
the biopsy (arrow’s direction)
2. Remove the needle-guided bracket:
a. Hold the prober in the left hand; unscrew the locking nut with the right hand to loose the
clamp (arrow’s direction).
b. Lift the biopsy up (towards arrow’s direction). The locating pole, front clamp, the locating
hole and the front clamp become loose.