11 Error messages
A dierentiation is made between the following errors:
- Dynamic errors are displayed with an error code (e.g. INF 01) for the duration of the
error.
- Temporary errors are displayed for 2 seconds (e.g. INF 07)
- Fatal errors are displayed permanently (e.g. ERR 101, can only be rectied via reset)
Display Cause Remedy
No display seg-
ments
There is no operating voltage Check electrical supply
------- Key has no function in this state
Flashing
Battery defective or time adjusted Set time
ERR 101 Key stuck
Key pressed when switching on
Release key or
consult Minebea Intec customer service
ERR 320 Operating program memory defective consult Minebea Intec customer service
ERR 340 New EEPROM loaded (service) Switch the scale o and on again; if
ERR 340 is permanently displayed, cont-
act Minebea Intec Customer Service
ERR 341 RAM lost data; battery empty Leave device switched on for at least
10 hours
ERR 343 Data loss in the memory for transaction number
external Alibi memory
consult Minebea Intec customer service
ERR 2627 The set time or the stored time of the Alibi me-
mory is implausible
Reset the time
INF 01 Data output does not t in the output format Make correct setting in the setup menu
INF 02 Calibration condition was not complied with, e.g.
not tared or load plate loaded
Only adjust after zero display, unload sca-
le, tare with key
INF 03 The calibration process could not be completed
within a certain time.
Allow the device to warm up and then re-
calibrate the device
INF 07 The last used function is not allowed for veried
scales
consult Minebea Intec customer service
INF 08 The load on the scale is too heavy to carry out ze-
roing
Please check whether the "power-on zero
range" (menu item [INIT.NUL] 1.12) was
complied with during the conguration.
INF 09 No tare possible at gross zero Zero the scales
INF 23 Error during initialization of an application consult Minebea Intec customer service
INF 29 Minimum load not reached Reduce minimum load (under application,
menu item [MIN.INIT] 3.6)
INF 71 Not possible to adopt the measured value (or the
entry) (e.g. control limit too small or too large)
None