EasyManua.ls Logo

Minebea Intec MiNexx-link-E PR 5215/01 - Page 102

Default Icon
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Operating | 6.14 - Calibrating internal weighing point | 102
6.14.2.4 Adjustment with weight
1. Click [by load] in the task bar.
2. Enter the weight value of the calibration weight in the line [Adjusted at].
3. Place the calibration weight on the scale.
4.
Click [Adjust] in the task bar.
Weight value, weight unit and measuring signal in mV/V
corresponding to this value are displayed in the line [Adjusted at]
5. Save the adjustment.
6.14.2.5 Adjusting with calculated mV/V value
The scale can be adjusted without weights. During input of the load cell mV/V
value, the acceleration of gravity at the place of installation can be taken into
account.
The PR- load cell data is based on the acceleration of gravity in Hamburg,
Germany:
9.81379m/s
2
.
Calculating Span
Span indicates the equivalent input voltage in mV/V related to the maximum
capacity (Max) of the scale. It is calculated as follows:
Span [mV/V]= max. capacity (Max) × load cell sensitivity C
n
 [mV/V]/ load cell
capacity (max. capacity E
max
× number of load cells)
load cell sensitivity C
n
= rated output C
n
(see technical data for the load cell).
Adjusting with mV/V
1. Click [by mV/V] in the task bar.
2. Enter the calculated Span value in the line [Adjusted at].
3.
Click [Adjust] in the task bar.
Weight value, weight unit and measuring signal in mV/V
corresponding to this value are displayed in the line [Adjusted at].
4. Save the adjustment.

Table of Contents