6. Operating | 6.14 - Calibrating internal weighing point | 103
6.14.2.6 Adjusting with load cell data (SMART calibration)
If the scale is not used in legal metrology, adjustment without weights can be
performed.
The easiest method is the one using load cell data without calculation. The
values are taken from the technical data and the calibration certificate of the
load cell.
1. Click in the task bar.
2. Enter the following values.
[Number of loadcells]
Number of load cells connected in parallel
Selection: 1 to 8
[Max. capacity of loadcell]
Maximum capacity E
max
of a load cell (not the total maximum capacity of
the scale!)
Input: For the value refer to the technical data of the load cell.
[Gravity]
Gravity at place of installation. Default is the value for Hamburg, Ger
many: 9.81379 m/s
2
.
[Hysteresis error]
Selection: not specified, specified
When switching from [not specified] to [specified] values for [Correction
A] and [Correction B] must be entered.
[Correction A]
Only with selection [specified]
Input: refer to the load cell certificate under "Hysteresis Correction A"
[Correction B]
Only with selection [specified]
Input: refer to the load cell certificate under "Hysteresis Correction B"
[Certified data]
Selection: All load cells the same, Each load cell individual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec MiNexx-link-E PR 5215/01 and is the answer not in the manual?

Minebea Intec MiNexx-link-E PR 5215/01 Specifications

General IconGeneral
BrandMinebea Intec
ModelMiNexx-link-E PR 5215/01
CategoryIndustrial Electrical
LanguageEnglish

Related product manuals