Signum 3 Ex
Operation
Operation
Weighing
Weighing W
The basic weighing function is always
accessible and can be used alone or in
combination with application programs,
such as Counting, Checkweighing,
Weighing in Percent, etc�
Characteristics
– Zeroing (
– Store the weight on the platform as tare
by pressing )
– Use the numeric keys to enter a tare
weight (press ) to save)
– Tare container weight automatically
– Delete tare values by entering 0
(press ) to save)
– Press k to toggle between:
– 1st and 2nd weight unit
– Display value and min� sample
quantity “SQmin"
Press K to toggle between:
– 10-fold increased resolution
(display max� 5 seconds)
Toggle L Net:
– Gross or net value
You can congure the k key function
in the Setup menu via:
Fn key
– Individual ID codes for weight values
– Print weight value
– Manually, by pressing p
– Automatically
(see “Data Output”)
– GMP printout
(see “Data Output”)
– Restore factory settings by selecting the
menu setting:
Appl: (Application)
Appl: def.app: 9.1
Automatic Taring
The first weight on the scale that
exceeds the preset minimum load is
stored in the tare memory at stability�
The values for subsequent loads are
stored as weight values� The scale
returns to the initial state when the
load is less than 50% of the minimum
load� Configure in Setup under:
Appl: (Application)
Appl a.tare
Minimum Load
To tare container weights automatically,
you need to set a minimum load in the
Setup menu, under:
Appl:(Application)
Appl: m.weigh
10 setting levels are available for
selection� They are dened in scale
intervals:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Example: if the scale interval is 1 g and
the minimum load is set to 1000 digits
(=1000 scale intervals), a load of at
least 1000 g is required for autotaring�
Automatic Printing
The rst weight value that exceeds the
minimum load is printed� Operating
menu setting:
setup:
prtprot: (Printout)
7.15. (Once at stability)
Device Parameters
Keypad
The keypad can be blocked�
Operating menu setting:
setup:
utilit: (Operating parameter)
8.3. (Keypad: blocking keys)
The following settings are available:
– 8.3.1. (All keys available)
– 8.3.2. (All keys blocked except e and
M)
– 8.3.3. (All alphanumeric keys blocked)
– 8.3.4 - 8.3.19 (Specied keys blocked
(see the menu under “Conguration”
for options))
Display
You can have the display backlighting
shut off automatically when not in use�
Operating menu setting:
setup:
utilit:
8.8. (Display lighting)
Automatic Shutdown
Operating menu setting:
setup:
utilit:
8.7. (Automatic shuto of indicator)
Timer
There are three timer settings:
two, four or ten minutes:
setup:
utilit:
8.9. (Timer)