34
Signum 3 Ex
Operation
Operation
Weighing
Weighing W
The basic weighing function is always
accessible and can be used alone or in
combination with application programs,
such as Counting, Checkweighing,
Weighing in Percent, etc
Characteristics
– Zeroing (
Store the weight on the platform as tare
by pressing )
Use the numeric keys to enter a tare
weight (press ) to save)
Tare container weight automatically
Delete tare values by entering 0
(press ) to save)
– Press k to toggle between:
– 1st and 2nd weight unit
Display value and min� sample
quantity “SQmin"
Press K to toggle between:
– 10-fold increased resolution
(display max� 5 seconds)
Toggle L Net:
– Gross or net value
You can congure the k key function
in the Setup menu via:
Fn key
Individual ID codes for weight values
Print weight value
– Manually, by pressing p
Automatically
(see “Data Output”)
GMP printout
(see “Data Output”)
Restore factory settings by selecting the
menu setting:
Appl: (Application)
Appl: def.app: 9.1
Automatic Taring
The first weight on the scale that
exceeds the preset minimum load is
stored in the tare memory at stability
The values for subsequent loads are
stored as weight values� The scale
returns to the initial state when the
load is less than 50% of the minimum
load� Configure in Setup under:
Appl: (Application)
Appl a.tare
Minimum Load
To tare container weights automatically,
you need to set a minimum load in the
Setup menu, under:
Appl:(Application)
Appl: m.weigh
10 setting levels are available for
selection� They are dened in scale
intervals:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Example: if the scale interval is 1 g and
the minimum load is set to 1000 digits
(=1000 scale intervals), a load of at
least 1000 g is required for autotaring�
Automatic Printing
The rst weight value that exceeds the
minimum load is printed� Operating
menu setting:
setup:
prtprot: (Printout)
7.15. (Once at stability)
Device Parameters
Keypad
The keypad can be blocked�
Operating menu setting:
setup:
utilit: (Operating parameter)
8.3. (Keypad: blocking keys)
The following settings are available:
8.3.1. (All keys available)
8.3.2. (All keys blocked except e and
M)
8.3.3. (All alphanumeric keys blocked)
8.3.4 - 8.3.19 (Specied keys blocked
(see the menu under “Conguration
for options))
Display
You can have the display backlighting
shut off automatically when not in use�
Operating menu setting:
setup:
utilit:
8.8. (Display lighting)
Automatic Shutdown
Operating menu setting:
setup:
utilit:
8.7. (Automatic shuto of indicator)
Timer
There are three timer settings:
two, four or ten minutes:
setup:
utilit:
8.9. (Timer)